dc.contributor.advisor | Di Donna, Paolo <1962> | |
dc.contributor.author | Catena, Andrea <2003> | |
dc.date.accessioned | 2025-10-16T14:30:01Z | |
dc.date.available | 2025-10-16T14:30:01Z | |
dc.date.issued | 2025-10-13 | |
dc.identifier.uri | https://unire.unige.it/handle/123456789/13203 | |
dc.description.abstract | Background: Nei pacemaker (PM) VDD monocatetere rappresentano un’ottima alternativa al PM bicamerale tradizionale. Tuttavia, la qualità del sensing atriale è cruciale per mantenere la sincronizzazione atrio-ventricolare (SAV). L’ipotesi è che un dipolo di sensing atriale lungo (>2 cm) assicuri un trascinamento AV più stabile rispetto a un dipolo corto (≤2 cm).
Obiettivo: Valutare, nella pratica clinica, l’efficacia del pacing VDD monocatetere nel preservare la SAV e il ruolo della lunghezza del dipolo atriale.
Metodi: Studio prospettico ambulatoriale in pazienti con impianto di PM VDD con valutazione clinica e strumentale impostando lower rate 50 bpm e sensibilità atriale <0.12 mV, con rivalutazione a 3 mesi. L’endpoint primario era la proporzione con SAV ≥90% tra le due coorti; endpoint secondario era la valutazione del sensing atriale e la correlazione con la %SAV. Analisi complete-case; confronto delle medie con t di Welch (α=0,05).
Risultati: Identificati 67 pazienti; esclusi 8 per dati mancanti e 8 per controindicazioni cliniche al 2° follow-up; coorte analitica 51 (25 dipolo lungo; 26 dipolo corto). SAV media ± DS: 95,4±4,6% (dipolo lungo) vs 84,6±25,5% (dipolo corto); mediana 97,0% vs 92,7%; Δ medie: +10,8%. Sensing atriale (mediana): 0,60 mV (dipolo lungo) vs 1,16 mV (dipolo corto).
Conclusioni: I cateteri VDD a dipolo lungo hanno mostrato un miglior trascinamento AV rispetto ai dipoli corti (~+11% di SAV). Nonostante il segnale atriale mediano più alto nel gruppo a dipolo corto, la maggiore stabilità/affidabilità del sensing con dipolo lungo ha garantito una migliore sincronizzazione AV. | it_IT |
dc.description.abstract | Background: In pacemakers (PMs), single-lead VDDs represent an excellent alternative to traditional dual-chamber PMs. However, the quality of atrial sensing is crucial for maintaining atrioventricular synchronization (AVS). The hypothesis is that a long atrial sensing dipole (>2 cm) ensures more stable AV tracking than a short dipole (≤2 cm).
Objective: To evaluate, in clinical practice, the effectiveness of single-lead VDD pacing in preserving AVS and the role of atrial dipole length.
Methods: Prospective outpatient study in patients with PM VDD implantation, with clinical and instrumental assessment using a lower rate of 50 bpm and atrial sensitivity <0.12 mV, with reassessment at 3 months. The primary endpoint was the proportion with AVS ≥90% between the two cohorts; secondary endpoint was the assessment of atrial sensing and its correlation with %AVS. Complete-case analysis; Comparison of means with Welch's t test (α = 0.05).
Results: 67 patients were identified; 8 were excluded due to missing data and 8 due to clinical contraindications at the second follow-up; analytic cohort: 51 (25 long dipole; 26 short dipole). Mean ± SD AVS: 95.4 ± 4.6% (long dipole) vs. 84.6 ± 25.5% (short dipole); median: 97.0% vs. 92.7%; mean Δ: +10.8%. Atrial sensing (median): 0.60 mV (long dipole) vs. 1.16 mV (short dipole).
Conclusions: Long dipole VDD catheters showed better AV tracking than short dipoles (~+11% AVS). Despite the higher median atrial signal in the short dipole group, the greater stability/reliability of long dipole sensing ensured better AV synchronization. | en_UK |
dc.language.iso | it | |
dc.language.iso | en | |
dc.rights | info:eu-repo/semantics/restrictedAccess | |
dc.title | Sincronizzazione atrio-ventricolare nel pacing VDD monocatetere: ruolo del dipolo di sensing atriale flottante | it_IT |
dc.title.alternative | Atrial- ventricular synchronisation in VDD pacing: role of the floating atrial sensing dipole | en_UK |
dc.type | info:eu-repo/semantics/bachelorThesis | |
dc.publisher.name | Università degli studi di Genova | |
dc.date.academicyear | 2024/2025 | |
dc.description.corsolaurea | 9291 - TECNICHE DI FISIOPATOLOGIA CARDIOCIRCOLATORIA E PERFUSIONE CARDIOVASCOLARE | |
dc.description.area | 6 - MEDICINA E CHIRURGIA | |
dc.description.department | 100007 - DIPARTIMENTO DI MEDICINA INTERNA E SPECIALITÀ MEDICHE | |